This is historical information of device classes implemented in AndorCamera device server.
Use this link to find the valid information.
Development status:
New development
Information status: Updated
Repository:
http://svn.code.sf.net/p/tango-ds/code/DeviceClasses/Acquisition/2D/AndorCamera
Contact:
Class Description
Class for connecting to and controlling the camera.
Families: Acquisition
Key words:
Platform: Unix Like
Language: Cpp
Contact:
Hardware
Manufacturer: Andor
Product:
CCD
Bus: USB
Class interface
Attributes:
| Name | Description |
|---|---|
| CoolingScalar: DevLong | Set on/off cooling. |
| CoolingTemperatureScalar: DevFloat | Writes desired temperature of the detector. Reads current temperature. |
| TriggerModeScalar: DevLong | Set trigger mode. Read last set value. |
| SpoolingScalar: DevLong | Activate/desactivate spooling. |
| FileDirScalar: DevString | Directory for saving files |
| SpoolingFileFormatScalar: DevLong | File Format for spooling files. Written to hardware only by writting to the Spooling attribute. |
| DataTypeScalar: DevLong | Writes the count convert mode. Reads the last set value. |
| AcquisitionModeScalar: DevLong | Set the acquisition mode. Reads the last set value. 1->singlescan,2->accumulate,3->Kinetics,4->FastKinetics,5->RunUntilAbort. |
| NumberAccumulationsScalar: DevLong | Set the number of scans accumulated in memory. It only takes effect in Accumulate or Kinetic Acquisition modes. Reads last set value. |
| ReadoutModeScalar: DevLong | Set the readout mode: 0-> full vertical binning, 1 -> multi track, 2 -> random track, 3 -> single track, 4 image |
| HBinScalar: DevLong | Last set horizontal binning |
| VBinScalar: DevLong | Last set vertical binning. |
| HStartScalar: DevLong | Last set horizontal RoI start. |
| HEndScalar: DevLong | Last set horizontal RoI end. |
| VStartScalar: DevLong | Last set vertical RoI start. |
| VEndScalar: DevLong | Last set vertical RoI end. |
| InternalShutterScalar: DevLong | Last internal shutter mode set. |
| ExternalShutterScalar: DevLong | Reads last set value. |
| TTLSignalShuttersScalar: DevLong | Reads the type of TTL signal send to the external shutter. |
| ExposureTimeScalar: DevDouble | Set exposure time and read the current valid value. |
| NumberKineticsScalar: DevLong | Set the number of scans taken in a single acquisition sequence. It only takes effect in Kinetic Acquisition mode. Reads last set value. |
| FilePrefixScalar: DevString | Prefix file name. Complete name: FilePrefix_FileNumber |
| FileNumberScalar: DevLong | File Number. Complete file name: FilePrefix_FileNumber |
| HShiftSpeedIDScalar: DevDouble | Set Horizontal Speed Shift by the index. Reads last set value. |
| HShiftSpeedScalar: DevFloat | Last set horizontal shift speed. |
| NbHShiftSpeedsScalar: DevLong | Number of horizontal shift speeds. |
| VShiftSpeedIDScalar: DevDouble | Set Verticalal Shift Speed by the index. Reads last set value. |
| VShiftSpeedScalar: DevFloat | Last set vertical shift speed. |
| NbVShiftSpeedsScalar: DevLong | Number of vertical shift speeds. |
Commands:
| Name | Description |
|---|---|
| StateInput: DevVoid Output: State Device state |
This command gets the device state (stored in its device_state data member) and returns it to the caller. |
| StatusInput: DevVoid Output: ConstDevString Device status |
This command gets the device status (stored in its device_status data member) and returns it to the caller. |
| SetImageBinRoIInput: DevVarLongArray HBin,VBin,HStart,HEnd,VStart,VEnd Output: DevVoid |
Set image bin und RoI parameters. |
| SetShutterInput: DevVarLongArray Int mode, ext mode, ttl signal Output: DevVoid |
Set internal and external shutter modes, and TTL signal. |
| GetHShiftSpeedByIndexInput: DevLong Shift Speed Index Output: DevFloat Shift Speed |
Reads the horizontal shift speed for the given index. |
| GetVShiftSpeedByIndexInput: DevLong Shift Speed Index Output: DevFloat Shift Speed |
Reads the vertical shift speed for the given index. |
| StartAcquisitionInput: DevVoid Output: DevVoid |
Start an acquisition. |
Pipes:
Properties:
| Name | Description |
|---|---|
| CameraIdDevLong | Id of the camera |
Updated:
The device class has been updated.
You are looking at this version now.
23 Feb 2017, Piotr Goryl
Updated:
The device class has been updated.
You can see previous version here .
19 Jan 2017, Piotr Goryl
Updated:
The device server has been updated.
You can see previous version here .
2 Jan 2017, Piotr Goryl
Created:
The device server has been added to catalogue.