Lambda2M - - TANGO Controls

This is historical information of device classes implemented in Lambda2M device server.

Use this link to find the valid information.

Development status: New development
Information status: Updated

Contact:



Class Description


This is a tango server project for lambda detetor.

Families: Acquisition

Key words:

Platform: Unix Like

Language: Cpp

License: GPL

Contact:

Hardware


Manufacturer: DESY(FS-DS)

Bus: Ethernet

Class interface


Attributes:

Name Description
TriggerModeScalar: DevShort TriggerMode of the Detector. 0: Internal trigger. ......
ShutterTimeScalar: DevDouble shutter time
DelayTimeScalar: DevDouble delay time, currently it is not used
FrameNumbersScalar: DevLong The frame numbers need to be acquired.
ThreadNoScalar: DevLong —
EnergyThresholdScalar: DevFloat set energy threshold
OperatingModeScalar: DevString operating mode of the detector: 24bit mode : TwentyFourBit 12bit mode : ContinuousReadWrite
ConfigFilePathScalar: DevString Path of the configuration file.
SaveAllImagesScalar: DevBoolean If this attribute is true, all data will be saved to disk
FilePrefixScalar: DevString The prefix of the created file. The full name is FilePrefix_FileStartNum.FilePostfix. Note if file splitting is enabled, the file name is like (to be decided).
FileStartNumScalar: DevLong The attribute is used when saving file. When file has been saved, the number is increased by one automatically. If user wants to overwrite the old file, the attribute value has to be reset. The full name is FilePrefix_FileStartNum.FilePostfix.
FilePreExtScalar: DevString file pre extention. If it is not empty, the final file name will be FilePrefix_FileStartNumFilePreExt.FilePostfix.
FilePostfixScalar: DevString Now the Lambda only writes file into hdf5 format, the postfix is h5. The full name is FilePrefix_FileStartNum.FilePostfix.
SaveFilePathScalar: DevString Save File Path. Define path for saving files.
SaveFileNameScalar: DevString define the file name for saving file.
LatestImageNumberScalar: DevLong The frame number of latest image
LiveModeScalar: DevBoolean live mode of the detector. Mainly used for online visualization. True: enable online visulization, frame rate is limited to 10fps. The data will not be saved to disk. The data can be fetched by accessing attribute Image False: diasble online visualization.
TotalLossFramesScalar: DevLong the total frame numbers lost during transmitting, used in 10GE linke mode.
CompressorShuffleScalar: DevBoolean during compression, shuffle the data or not
CompressionRateScalar: DevUShort set compression rate from 0-9
CompressionEnabledScalar: DevBoolean applying compression to data when saving. True: compression False: no compression
LayoutScalar: DevString The layout out of Chips. 1 : 1536x512 pixels 2 : 768x512 pixels
ShutterTimeMaxScalar: DevDouble define the maximum shutter time(only for expert view).
ShutterTimeMinScalar: DevDouble define the minimum shutter time(only for expert view).
WidthScalar: DevULong Image width
HeightScalar: DevULong Image height
DepthScalar: DevULong Pixel depth.
LiveFrameNoScalar: DevLong Current frame No. which is related to the Image attribute
DistortionCorrectionScalar: DevUShort distortion correction 1 : division; Large pixels are divided according to predefined denominator. The value of the pixel are rounded during division. If pixel value is saturated, the division is not applied.
CreateDirScalar: DevBoolean if the attribute is true, directory is created when it doesn`t exist.
AppendDataScalar: DevBoolean append data to old file. This is useful when it only takes one image at each scan point during scan.
CheckFileExistsScalar: DevBoolean —
PrecompressEnabledScalar: DevBoolean —
FramesPerFileScalar: DevLong —
FileNamePathCaseSensitiveScalar: DevBoolean —
OpModeScalar: DevLong —
LiveLastImageDataImage: DevLong provide last image data for the live mode

Commands:

Name Description
StateInput: DevVoid
Output: State
Device state
This command gets the device state (stored in its device_state data member) and returns it to the caller.
StatusInput: DevVoid
Output: ConstDevString
Device status
This command gets the device status (stored in its device_status data member) and returns it to the caller.
StartAcqInput: DevVoid
Output: DevVoid
Start the acquisition.
StopAcqInput: DevVoid
Output: DevVoid
stop the acquistion

Pipes:

Properties:

Name Description
SimulationModeDevUShort Providing simulation mode for the detector
ConfigurationPathDevString path of configuration file
SubDevicesArray of DevString —

22 May 2019, DS Admin
Updated:
The device class has been updated.
You can see previous version here .



2 Mar 2019, DS Admin
Updated:
The device class has been updated.
You can see previous version here .



23 Feb 2019, DS Admin
Updated:
The device class has been updated.
You can see previous version here .



2 Feb 2019, DS Admin
Updated:
The device class has been updated.
You can see previous version here .



8 Dec 2018, DS Admin
Updated:
The device class has been updated.
You are looking at this version now.



28 Jul 2018, DS Admin
Updated:
The device class has been updated.
You can see previous version here .



7 Jul 2018, DS Admin
Updated:
The device class has been updated.
You can see previous version here .



23 Feb 2017, Piotr Goryl
Updated:
The device class has been updated.
You can see previous version here .



19 Jan 2017, Piotr Goryl
Updated:
The device server has been updated.
You can see previous version here .



2 Jan 2017, Piotr Goryl
Created:
The device server has been added to catalogue.