This is historical information of device classes implemented in DigitalDelayGenerator device server.
Use this link to find the valid information.
Development status:
Released
Information status: New
Contact:
Class Description
- This device controls a T560 Digital Delay Generator :
- four TTL-level pulse outputs, each programmable for delay and width up to 10 seconds each, with 10-picosecond resolution.
- Programmable-level trigger input with divide/burst features and trigger enable GATE input.
Controlled via a RS232 serial interface or a 10/100-mbps Ethernet interface.
NOTE : - RS232 config : 38400, 8, N, 1 - ETHERNET port : 2000
Families: CounterTimer
Key words: Instrumentation
Platform: All Platforms
Language: Cpp
Contact:
Hardware
Manufacturer: Highland Technology
Product:
T560
Bus: Ethernet
Class interface
Attributes:
| Name | Description |
|---|---|
| shotsCounterScalar: DevULong | — |
| pulseDelayChannelAScalar: DevDouble | — |
| pulseWidthChannelAScalar: DevDouble | — |
| activateChannelAScalar: DevBoolean | — |
| pulseDelayChannelBScalar: DevDouble | — |
| pulseWidthChannelBScalar: DevDouble | — |
| activateChannelBScalar: DevBoolean | — |
| pulseDelayChannelCScalar: DevDouble | — |
| pulseWidthChannelCScalar: DevDouble | — |
| activateChannelCScalar: DevBoolean | — |
| pulseDelayChannelDScalar: DevDouble | — |
| pulseWidthChannelDScalar: DevDouble | — |
| activateChannelDScalar: DevBoolean | — |
Commands:
| Name | Description |
|---|---|
| StateInput: DevVoid Output: State |
This command gets the device state (stored in its device_state data member) and returns it to the caller. |
| StatusInput: DevVoid Output: DevString |
This command gets the device status (stored in its device_status data member) and returns it to the caller. |
| HardwareResetInput: DevVoid Output: DevVoid |
Performs a hardware reset/restart of the T560 (equivalent to a power off/on cycle. NOTE : the reset takes about 4 seconds, after which the T560 will respond with the following string "Highland Thechnology T560 DDG " The last-saved setup will be installed. |
| ClearErrorsInput: DevVoid Output: DevVoid |
Clear error flag. |
| ResetShotsInput: DevVoid Output: DevVoid |
Reset shot counter. |
| SaveModuleConfigurationInput: DevVoid Output: DevVoid |
Save the T560 module configuration in its intrenal flash memory. This configuartion is applied after a reset. |
| RecallModuleConfigurationInput: DevVoid Output: DevVoid |
Reapply the previous module configuration. No need to do a reset. |
Pipes:
Properties:
| Name | Description |
|---|---|
| CommunicationProxyDevString | Name of the device which controls the communication link. This module support both RS232 and Ethernet interfaces. |
Updated:
The device class has been updated.
You can see previous version here .
20 Apr 2017, Piotr Goryl
Updated:
The device class has been updated.
You are looking at this version now.
23 Feb 2017, Piotr Goryl
Created:
The device class has been added to catalogue.