This is historical information of device classes implemented in Mythen device server.

Use this link to find the valid information.

Development status: Released
Information status: Updated

Contact:



Class Description


Device server for the Mythen detector.

Families: Acquisition

Key words:

Platform: Unix Like

Language: Cpp

License:

Contact:

Hardware


Manufacturer: PSI

Product:

Mythen

Bus: Data Socket

Class interface


Attributes:

Name Description
OnlineScalar: DevShort Detector is Online (1) or Offline (0).
ConfigFileScalar: DevString The name of the configuration file.
ParameterFileScalar: DevString File containing the main detector parameters for the\nmeasurement.
SetupFileScalar: DevString This is the root of the filename for the complete detector\nsetup (including configuration, trimbits, flat field\ncoefficients, badchannels etc.). The extension is\nautomatically generated.
CalibrationDirScalar: DevString The path of the calibration files.
TrimDirScalar: DevString The path of the trim files.
TrimEnergiesScalar: DevString The number of energies for which trimbit files exist\nand their value.\nString format: nen [e0 e1...en]
OutDirScalar: DevString The directory to which the files will be written by default.
OutFileNameScalar: DevString The filename to which the files will be written by default\n(to which file and position indexes will eventually be\nattached).
OutFileIndexScalar: DevLong The start index of the output files (automatically\nincremented by the acquisition functions).
EnableMergeScalar: DevBoolean Set to 'true' to enable merging of frames to a single image. Otherwise\nthey will be kept in OutDir.
MinFramesScalar: DevShort If less than MinFrames frames are acquired, they are not merged in a\nsingle image, but just moved to MergeDir.
MergeDirScalar: DevString The directory to which the frames will be merged (or just moved).
MergeOngoingScalar: DevBoolean This flag is 'true' when a merging is ongoing.
NMaxModulesScalar: DevShort The maximum number of detector modules.
NModulesScalar: DevShort
SettingsScalar: DevString Detector settings. Can be:\n uninitialised\n standard\n fast\n highgain
ThresholdScalar: DevShort Detector threshold in eV.\nShould be half of the beam energy.\nIt is precise only if the detector is calibrated.\n
ExposureTimeScalar: DevFloat The exposure time per frame (in s).
FramePeriodScalar: DevFloat The frame period (in s).
DelayAfterTrgScalar: DevFloat The delay after trigger (in s).
GatesPerFrameScalar: DevShort The number of gates per frame.
FramesPerCycleScalar: DevLong The number of frames per cycle (e.g. after each trigger).
NCyclesScalar: DevShort The number of cycles (e.g. number of triggers).
NProbesScalar: DevShort The number of probes to accumulate (max 3).
DynamicRangeScalar: DevShort The dynamic range - can be 1, 4, 8,16 or 24 bits.
FlagsScalar: DevString Readout flags - can be none or storeinram.
ThreadedScalar: DevShort This tells whether the postprocessing and file writing of\nthe data is done in a separate thread (0 sequencial,\n1 threaded). Check that it is set to 1 if you acquire\nlong real time measurements and/or use the storeinram\noption otherwise you risk to lose your data.
TrimModeScalar: DevString Trims the detector and writes the trimfile.\nMode can be:noise, beam, improve, fix, offline.\nCheck that the start conditions are OK!!!
TrimFileScalar: DevString Name of trimfile (without extension .snxx).\nCheck that the start conditions are OK!!!
FlatFieldEnabledScalar: DevBoolean True when the flat field corrections are enabled.
FlatFieldFileNameScalar: DevString File containing the flat field corrections.
RateCorrEnabledScalar: DevBoolean True when the rate corrections are enabled.
RateCorrDefaultScalar: DevBoolean True when using default rate corrections.
RateCorrDeadTimeScalar: DevFloat The dead time used for rate corrections.
BadChannelsEnabledScalar: DevBoolean True when the the bad channels corrections are enabled.
BadChannelsFileNameScalar: DevString File containing the definition of the bad channels.
AngConvEnabledScalar: DevBoolean True when the the angular conversion is enabled.
AngConvFileNameScalar: DevString File containing the definition of the angular conversion\ncoefficients.
GlobalOffsetScalar: DevFloat The fixed angular offset of your encoder -\nshould be almost constant!
FineOffsetScalar: DevFloat A possible angular offset of your setup -\nshould be small but can be senseful to modify.
BinSizeScalar: DevFloat The binning size of the angular conversion.
PositionsScalar: DevString The number of positions at which the detector is moved\nduring the acquisition and their values.\nString format: np [pos0 pos1...posnp]
ModuleSerialNumberScalar: DevString The module serial number(s).
ModuleFirmwareVersionScalar: DevLong64 The module firmware version.
DetectorSerialNumberScalar: DevLong64 The detector serial number (MAC address).
DetectorFirmwareVersionScalar: DevLong64 The detector firmware version.
DetectorSoftwareVersionScalar: DevLong64 The detector software version.
MythenSoftwareVersionScalar: DevLong64 The mythen software version.
DigitalTestResultScalar: DevString Result of the Digital Test(s):\n0 test passed
BusTestResultScalar: DevString Result of the Bus Test:\n0 test passed\n
ExtSigResultScalar: DevString Result of the ExtSig Command.\n
DataSpectrum: DevULong Last image data.

Commands:

Name Description
StateInput: DevVoid
Output: State
State Code
This command gets the device state (stored in its <i>device_state</i> data member) and returns it to the caller.
StatusInput: DevVoid
Output: ConstDevString
Status description
This command gets the device status (stored in its <i>device_status</i> data member) and returns it to the caller.
RefreshInput: DevVoid
Output: DevVoid
Read all values and configurations, and update the values shown to the user.
StartAcquisitionInput: DevVoid
Output: DevVoid
Start the acquisition.
StopAcquisitionInput: DevVoid
Output: DevVoid
Stop the acquisition.
ResetInput: DevVoid
Output: DevVoid
Restart the mythen host.
GetDataInput: DevVoid
Output: DevVoid
Get all data from the detector (if any), process them and write them to file according to the preferences already setup. In principle the method is not needed (when using StartAcquisition the files are written automatically to disk).
GetFrameInput: DevVoid
Output: DevVoid
Get a single frame from the detector (if any), processe it and write it to file according to the preferences already setup. In principle the method is not needed (when using StartAcquisition the files are written automatically to disk).
SetConfigInput: DevVoid
Output: DevVoid
Read the configuration file and set the values.
GetConfigInput: DevVoid
Output: DevVoid
Write the configurations to file.
SetParametersInput: DevVoid
Output: DevVoid
Read the detector parameters from file and set them.
GetParametersInput: DevVoid
Output: DevVoid
Write the main detector parameters to file.
ReadSetupFromFileInput: DevVoid
Output: DevVoid
Read the files specified (and that could be created by get setup) and reset the complete detector configuration including flatfield corrections, badchannels, trimbits etc.
SaveSetupToFileInput: DevVoid
Output: DevVoid
Writes the complete detector setup (including configuration, trimbits, flat field coefficients, Write the complete detector setup (including configuration, trimbits, flat field coefficients, badchannels etc.) in a set of files for which the extension is automatically generated.
ExtSigGetInput: DevShort
Index of the external digital signal
Output: DevString
Usage of the external digital signal
Get usage of the external digital signal i.
ExtSigSetInput: DevString
Index of the external digital signal and mode (eg: 1 off)
Output: DevString
Usage of the external digital signal
Set usage of the external digital signal i.
RunDigitalTestInput: DevVoid
Output: DevVoid
Make a digital test of the detector. Return 0 if it succeeds.
RunBusTestInput: DevVoid
Output: DevVoid
Make a test of the detector bus. Return 0 if it succeeds.
WriteTrimFileInput: DevVoid
Output: DevVoid
Trim the detector and write the trimfile.
GetFlatFieldCorrectionsInput: DevVoid
Output: DevVoid
Write the flat field corrections to file.
GetBadChannelsInput: DevVoid
Output: DevVoid
Write the bad channels to file.
GetAngularConversionsInput: DevVoid
Output: DevVoid
Write the angular conversion coefficients to file.

Pipes:

Properties:

Name Description
SimulationModeDevShort 0 real mode, 1 simulation mode.
HostnameDevString
PortDevShort The detector port.
DetectorIdDevShort The detector ID.

23 Feb 2017, Piotr Goryl
Updated:
A device class has been added.
You can see previous version here .



23 Feb 2017, Piotr Goryl
Updated:
The device class has been updated.
You can see previous version here .



19 Jan 2017, Piotr Goryl
Updated:
The device server has been updated.
You can see previous version here .



19 Jan 2017, Piotr Goryl
Updated:
The device server has been updated.
You are looking at this version now.



2 Jan 2017, Piotr Goryl
Updated:
The device server has been updated.
You can see previous version here .



2 Jan 2017, Piotr Goryl
Created:
The device server has been added to catalogue.