Average: 0
Rating Count: 0

Development status: New development
Information status: New
Repository: http://svn.code.sf.net/p/tango-ds/code/DeviceClasses/MeasureInstruments/GEpressCtrlPACE5000Channel
Contact:



Class Description


Server for the controll of the Pressuer Automated Calibration Equipment 5000 from GE

Families: MeasureInstruments

Key words:

Platform: Unix Like

Language: Python

License: GPL

Contact:

Hardware


Manufacturer: GE

Product:

PACE-5000

Bus: Socket

Class interface


Attributes:

Name Description
DecompressRateScalar: DevDouble Decompression rate set by the user. Used in conjunction with the Decompress Tango Device Command
SourceEffectivenessScalar: DevDouble Parameter reflecting effectiveness of the effort in order to achieve the setpoint
PressureControlScalar: DevBoolean Parameter controlling the main operation of the pump. If pressure control is ON – the device does it best to increase or decrease the pressure according to the setpoint. f pressure control is OFF – the device is going to the measurement mode, no control, just measurement (useful for leakage testing)
OvershootScalar: DevBoolean Parameter controlling the overshooting capability on pressure increase: 0 – not allowed 1 – allowed
SlewRateModeScalar: DevBoolean Instrumental mode for the slew rate operation – controls usage of the SlewRate parameter. Maximum slew rate – SlewRate parameter is disregarded and compression or decompression takes place with maximum pace. Linear – device respects the SlewRate setting and performs the compression or decompression accordingly
SlewRateScalar: DevDouble Slew rate for the output line pressure control
SetPointScalar: DevDouble Setpoint output line pressure
InletPressureScalar: DevDouble Current input pressure
MembranePressureScalar: DevDouble Current output pressure

Commands:

Name Description
CompressInput: DevVoid
Output: DevVoid
Starts the pump to do compression or decompression PressureControl = 1
DecompressInput: DevVoid
Output: DevVoid
Starts the decompression to the lowest pressure with DecompressRate First – stops the operation of the pump. Reads the value of the DecompressRate, sets the value of the SlewRate to this value Sets the SlewRateMode to 0=LINEAR Sets the SetPoint = 0 Sets the PressureControl = 1
StopInput: DevVoid
Output: DevVoid
Stops operation of the pump. The most important action – stop the PressureControl – power control – go to the measurement mode
StatusInput: DevVoid
Output: ConstDevString
Device status
This command gets the device status (stored in its device_status data member) and returns it to the caller.
StateInput: DevVoid
Output: State
Device state
This command gets the device state (stored in its device_state data member) and returns it to the caller.

Pipes:

Properties:

Name Description
ChannelDevLong Channel number (1 or 2)
SocketDeviceDevString Name of the device of the GEpressCtrlPACE5000 class making the socket connection


Please log in to comment.

2 Feb 2019, DS Admin
Created:
The device class has been added to catalogue.
Added by:dadmin on:2 Feb 2019, 4:34 a.m.